{"id":21691,"date":"2023-06-05T13:37:53","date_gmt":"2023-06-05T05:37:53","guid":{"rendered":"https:\/\/www.bihec.com\/additel\/?p=21691"},"modified":"2023-06-05T13:37:53","modified_gmt":"2023-06-05T05:37:53","slug":"additel-760%e6%89%8b%e6%8c%81%e5%bc%8f%e8%87%aa%e5%8a%a8%e5%8e%8b%e5%8a%9b%e6%a0%a1%e5%87%86%e5%99%a8","status":"publish","type":"post","link":"https:\/\/www.bihec.com\/additel\/additel-760%e6%89%8b%e6%8c%81%e5%bc%8f%e8%87%aa%e5%8a%a8%e5%8e%8b%e5%8a%9b%e6%a0%a1%e5%87%86%e5%99%a8\/","title":{"rendered":"Additel 760\u624b\u6301\u5f0f\u81ea\u52a8\u538b\u529b\u6821\u51c6\u5668"},"content":{"rendered":"\n\n\n
\n

\"\"<\/p>\n<\/td>\n

\u5e26\u5185\u7f6e\u6cf5\u548c\u63a7\u5236\u5668\u7684\u5168\u81ea\u52a8\u6821\u51c6\u5668<\/p>\n

\u53ef\u5207\u6362\u7684\u5185\u90e8\u538b\u529b\u6a21\u5757\uff0c\u53ef\u6269\u5c55\u8303\u56f4<\/p>\n

\u7cbe\u5ea6\uff081\u5e74\uff090.02%FS<\/p>\n

\u5916\u90e8\u538b\u529b\u6a21\u5757\u53ef\u7528\uff08\u4ec5\u9650\u6d4b\u91cf\uff09<\/p>\n

\u624b\u6301\u64cd\u4f5c\u5c0f\u4e8e4\u78c5\uff081.8\u516c\u65a4\uff09<\/p>\n

\u6e90\u538b\u529b\u3001\u6d4b\u91cf\u538b\u529b\u548c\u7535\u6c14<\/p>\n

4\u4e2a\u901a\u9053<\/p>\n

\u53ef\u9009HART\u901a\u4fe1<\/p>\n

\u53ef\u9009\u7684\u6570\u636e\u8bb0\u5f55\u548c\u4efb\u52a1\u6587\u6863<\/p>\n

USB\u548cWi-Fi\u901a\u4fe1<\/p>\n

ISO 17025\u8ba4\u8bc1\u6821\u51c6\uff0c\u5305\u62ec\u6570\u636e<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n

\u4e00\u4e2a\u4fbf\u643a\u5f0f\u81ea\u52a8\u538b\u529b\u6821\u51c6\u5668\u5728\u4f60\u7684\u624b\u638c\u5fc3\u8fd9\u53ef\u80fd\u662f\u6211\u4eec\u8fc4\u4eca\u4e3a\u6b62\u6700\u4ee4\u4eba\u5174\u594b\u7684\u4ea7\u54c1\uff01Additel 760\u7cfb\u5217\u624b\u6301\u5f0f\u81ea\u52a8\u538b\u529b\u6821\u51c6\u5668\u5c06\u4fbf\u643a\u5f0f\u538b\u529b\u6821\u51c6\u63d0\u5347\u5230\u65b0\u7684\u6c34\u5e73\u3002ADT760\u7684\u521b\u65b0\u8bbe\u8ba1\u91cd\u91cf\u4e0d\u52304\u78c5\uff081.8\u516c\u65a4\uff09\uff0c\u5305\u542b\u5185\u7f6e\u6cf5\u3001\u7cbe\u5bc6\u538b\u529b\u4f20\u611f\u5668\u3001\u5185\u90e8\u63a7\u5236\u5668\u548c\u5927\u578b\u89e6\u6478\u5c4f\u5f69\u8272\u663e\u793a\u5c4f\u3002\u8981\u4ea7\u751f\u538b\u529b\uff0c\u53ea\u9700\u8f93\u5165\u6240\u9700\u538b\u529b\uff0cAdditel 760\u5373\u53ef\u5b8c\u6210\u5176\u4f59\u64cd\u4f5c\u3002\u6bcf\u4e2a\u5355\u5143\u6709\u56db\u4e2a\u901a\u9053\uff1a\u4e00\u4e2a\u7528\u4e8e\u6e90\u548c\u6d4b\u91cf\u538b\u529b\u7684\u5185\u90e8\u538b\u529b\u901a\u9053\u3001\u4e24\u4e2a\u5916\u90e8\u538b\u529b\u6d4b\u91cf\u901a\u9053\u4ee5\u53ca\u4e00\u4e2a\u7535\u5b50\u6d4b\u91cf\u548c\u6e90\u901a\u9053\u3002\u8be5\u7cfb\u5217\u6821\u51c6\u5668\u6709\u4e09\u79cd\u6807\u51c6\u578b\u53f7\uff0c\u53ef\u9009\u62e9\u6dfb\u52a0HART\u901a\u4fe1\u3001\u6587\u4ef6\u548c\u6570\u636e\u8bb0\u5f55\u3002<\/p>\n

\u578b\u53f7<\/strong><\/p>\n

ADT760-LLP\uff08\u7a7a\u6c14\uff1a\u9ad8\u8fbe\u00b130 inH2O\uff09<\/strong><\/p>\n

\u4e13\u4e3a\u4f4e\u538b\u6821\u51c6\u800c\u8bbe\u8ba1\uff0c\u5e76\u914d\u6709\u60a8\u9009\u62e9\u7684\u5185\u7f6e\u538b\u529b\u6a21\u5757\u3002\u4e0eADT760-LLP\u517c\u5bb9\u7684\u6700\u5927\u91cf\u7a0b\u6a21\u5757\u4e3a\u00b130 inH2O\uff08\u00b175 mbar\uff09\uff0c\u5e76\u63d0\u4f9b0.05%FS\u7684\u7cbe\u5ea6\uff08\u6709\u5173ADT760\u9009\u9879\u548c\u60a8\u9009\u62e9\u7684\u6a21\u5757\u7684\u914d\u7f6e\uff0c\u8bf7\u53c2\u9605\u8ba2\u8d2d\u4fe1\u606f\uff09\u3002\u63d0\u4f9b\u989d\u5916\u7684\u5185\u90e8\u538b\u529b\u6a21\u5757\uff08ADT155\u7cfb\u5217\uff09\uff0c\u53ef\u63d0\u4f9b\u4f4e\u81f3\u00b10.25 inH2O\uff08\u00b10.62 mbar\uff09\u7684\u5404\u79cd\u8303\u56f4\u30020.05%FS\u7684\u7cbe\u5ea6\u548c0.005%FS\u7684\u63a7\u5236\u7a33\u5b9a\u6027\u57fa\u4e8e\u5185\u90e8\u6a21\u5757\u7684\u8de8\u5ea6\u3002\u6d4b\u91cf\u53ef\u4ee5\u5728\u4eea\u8868\u6216\u5dee\u52a8\u6a21\u5f0f\u4e0b\u8fdb\u884c\u3002<\/p>\n

ADT760-D\uff08\u7a7a\u6c14\uff1a\u6700\u9ad825 psi\uff09<\/strong><\/p>\n

\u4e3a\u60a8\u63d0\u4f9b\u5dee\u538b\u548c\u8868\u538b\uff0c\u4f46\u538b\u529b\u8303\u56f4\u9ad8\u4e8eADT760-LLP\u3002ADT760-D\u7684\u91cf\u7a0b\u4e3a-12.5\u81f335\u78c5\/\u5e73\u65b9\u82f1\u5bf8\uff08-0.86\u81f32.5\u5df4\uff09\uff0c\u7cbe\u5ea6\u4e3a0.02%FS\uff0c\u662f\u8986\u76d6\u975e\u5e38\u5e38\u89c1\u7684\u8868\u538b\u548c\u5dee\u538b\u6d4b\u91cf\u7684\u7406\u60f3\u89e3\u51b3\u65b9\u6848\u3002Additel 760-D\u914d\u5907\u4e86\u4e00\u4e2a\u8986\u76d6\u5176\u5168\u91cf\u7a0b\u7684\u5185\u90e8\u6a21\u5757\uff0c\u4f46\u53ef\u4ee5\u8d2d\u4e70\u4f4e\u81f3\u00b110 inH2O\u5dee\u538b\u7684\u4f4e\u538b\u914d\u7f6e\uff0c\u4ee5\u63d0\u9ad8\u4f4e\u538b\u4e0b\u7684\u7cbe\u5ea6\u3002<\/p>\n

ADT760-MA\uff08\u7a7a\u6c14\uff1a\u6700\u9ad8300 psig\uff09<\/strong><\/p>\n

760-MA\u53ef\u4ea7\u751f\u548c\u63a7\u5236\u9ad8\u8fbe300 psig\uff0820 bar.g\uff09\u7684\u771f\u7a7a\u538b\u529b\uff0c\u7cbe\u5ea6\u4e3a0.02%FS\u3002\u914d\u6709\u5185\u7f6e\u6c14\u538b\u57fa\u51c6\uff0c\u6bcf\u4e2a\u88c5\u7f6e\u53ef\u5728\u8868\u538b\u548c\u7edd\u5bf9\u538b\u529b\u7c7b\u578b\u4e4b\u95f4\u5207\u6362\u3002\u53ef\u63d0\u4f9b\u5404\u79cd\u5185\u90e8\u4f20\u611f\u5668\uff0c\u8fd9\u4e9b\u4f20\u611f\u5668\u63d0\u4f9b\u8f83\u4f4e\u7684\u538b\u529b\u8303\u56f4\u4ee5\u63d0\u9ad8\u6027\u80fd\u3002<\/p>\n

\n
\n\n\n\n\n\n\n\n\n\n\n\n\n\n
Pressure Specification<\/strong><\/th>\n<\/tr>\n<\/thead>\n
Max Pressure Range<\/strong><\/td>\n-12.5 to 35 psi (-0.86 to 2.5 bar)<\/td>\n<\/tr>\n
Accuracy<\/strong><\/td>\n0.02% FS<\/td>\n<\/tr>\n
Stability<\/strong><\/td>\n<0.005% FS<\/td>\n<\/tr>\n
Pressure Type<\/strong><\/td>\nDifferential, Gauge<\/td>\n<\/tr>\n
Over Range Indication<\/strong><\/td>\n1.2<\/td>\n<\/tr>\n
Resolution<\/strong><\/td>\n6 digits<\/td>\n<\/tr>\n
Measurement Units<\/strong><\/td>\nPa, hPa, kPa, mPa, bar, mbar, msi, mmHg @ 0\u00b0C, cmHg @ 0\u00b0C, mHg @ 0\u00b0C, inHg @ 0\u00b0C, inH2O @ 4\u00b0C, mmH2O @ 4\u00b0C, cmH2O @ 4\u00b0C, mH2O @ 4\u00b0C, inH2O @ 20\u00b0C, mmH2O @ 20\u00b0C, cmH2O @ 20\u00b0C, mH2O @ 20\u00b0C, inH2O @ 68\u00b0C, kgf\/cm2, mtorr, torr, lb\/ft2, tsi, custom<\/td>\n<\/tr>\n
Connection<\/strong><\/td>\nHose, 5 ft (1.5 m), with built-in filter to 1\/4BSPF, 1\/4NPTF, and M20F adapters<\/td>\n<\/tr>\n
Pressure Output Rate<\/strong><\/td>\n<10 Seconds (35 psi\/5 ml)<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n

 <\/p>\n<\/div>\n

\n
\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n
Selection Guide – Differential Pressure<\/strong><\/th>\n<\/tr>\n
Module<\/strong><\/th>\nModule Range[6]<\/strong><\/th>\nMedia<\/strong><\/th>\nAccuracy (%FS)[1]<\/strong><\/th>\nBurst Pressure<\/strong><\/th>\n760-LLP<\/strong><\/th>\n760-D<\/strong><\/th>\n760-MA<\/strong><\/th>\n<\/tr>\n
inH2O<\/strong><\/th>\nmbar<\/strong><\/th>\n<\/tr>\n<\/thead>\n
DP025<\/strong><\/td>\n\u00b10.25<\/td>\n\u00b10.62<\/td>\nG<\/td>\n0.2[2]<\/td>\n100x<\/td>\n\u2022<\/td>\n<\/td>\n<\/td>\n<\/tr>\n
DP050<\/strong><\/td>\n\u00b10.5<\/td>\n\u00b11.25<\/td>\nG<\/td>\n0.1[3]<\/td>\n100x<\/td>\n\u2022<\/td>\n<\/td>\n<\/td>\n<\/tr>\n
DP1<\/strong><\/td>\n\u00b11<\/td>\n\u00b12.5<\/td>\nG<\/td>\n0.05[4]<\/td>\n100x<\/td>\n\u2022<\/td>\n<\/td>\n<\/td>\n<\/tr>\n
DP2<\/strong><\/td>\n\u00b12<\/td>\n\u00b15<\/td>\nG<\/td>\n0.05[4]<\/td>\n100x<\/td>\n\u2022<\/td>\n<\/td>\n<\/td>\n<\/tr>\n
DP5<\/strong><\/td>\n\u00b15<\/td>\n\u00b110<\/td>\nG<\/td>\n0.05[4]<\/td>\n50x<\/td>\n\u2022<\/td>\n<\/td>\n<\/td>\n<\/tr>\n
DP10<\/strong><\/td>\n\u00b110<\/td>\n\u00b125<\/td>\nG<\/td>\n0.05[4]<\/td>\n20x<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP20<\/strong><\/td>\n\u00b120<\/td>\n\u00b150<\/td>\nG<\/td>\n0.05<\/td>\n20x<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP30<\/strong><\/td>\n\u00b130<\/td>\n\u00b175<\/td>\nG<\/td>\n0.05<\/td>\n20x<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP50<\/strong><\/td>\n\u00b150<\/td>\n\u00b1125<\/td>\nG<\/td>\n0.05<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP100<\/strong><\/td>\n\u00b1100<\/td>\n\u00b1250<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP150<\/strong><\/td>\n\u00b1150<\/td>\n\u00b1350<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP300<\/strong><\/td>\n\u00b1300<\/td>\n\u00b1700<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP400<\/strong><\/td>\n-380 to 400<\/td>\n-950 to 1K<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP800<\/strong><\/td>\n-380 to 800<\/td>\n-950 to 2K<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
DP1K<\/strong><\/td>\n-380 to 1K<\/td>\n-950 to 2.5K<\/td>\nG<\/td>\n0.02<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n<\/td>\n<\/tr>\n
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.<\/td>\n<\/tr>\n
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS.<\/td>\n<\/tr>\n
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS<\/td>\n<\/tr>\n
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS.<\/td>\n<\/tr>\n
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA<\/td>\n<\/tr>\n
[6] The low module pressure range may be outside the pressure range of the calibrator<\/td>\n<\/tr>\n
* ADT155 Pressure Modules are calibrated in psi and inH2O.<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n

 <\/p>\n<\/div>\n

\n
\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n
Selection Guide – Gauge Pressure<\/strong><\/th>\n<\/tr>\n
Module<\/strong><\/th>\nModule Range[6]<\/strong><\/th>\nMedia<\/strong><\/th>\nAccuracy (%FS)[1]<\/strong><\/th>\nBurst Pressure<\/strong><\/th>\n760-LLP<\/strong><\/th>\n760-D<\/strong><\/th>\n760-MA<\/strong><\/th>\n<\/tr>\n
psi<\/strong><\/th>\nbar<\/strong><\/th>\n<\/tr>\n<\/thead>\n
CP10<\/strong><\/td>\n\u00b110<\/td>\n\u00b10.7<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/tr>\n
CP15<\/strong><\/td>\n-13.5 to 15<\/td>\n-0.95 to 1<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/tr>\n
CP30<\/strong><\/td>\n-13.5 to 30<\/td>\n-0.95 to 2<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/tr>\n
CP35<\/strong><\/td>\n-13.5 to 35<\/td>\n-0.95 to 2.5<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n\u2022<\/td>\n\u2022<\/td>\n<\/tr>\n
CP50<\/strong><\/td>\n-13.5 to 50<\/td>\n-0.95 to 3.5<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n<\/td>\n\u2022<\/td>\n<\/tr>\n
CP100<\/strong><\/td>\n-13.5 to 100<\/td>\n-0.95 to 7<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n<\/td>\n\u2022<\/td>\n<\/tr>\n
CP150<\/strong><\/td>\n-13.5 to 150<\/td>\n-0.95 to 10<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n<\/td>\n\u2022<\/td>\n<\/tr>\n
CP200<\/strong><\/td>\n-13.5 to 200<\/td>\n-0.95 to 14<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n<\/td>\n\u2022<\/td>\n<\/tr>\n
CP300<\/strong><\/td>\n-13.5 to 300<\/td>\n-0.95 to 20<\/td>\nG<\/td>\n0.02[5]<\/td>\n3x<\/td>\n<\/td>\n<\/td>\n\u2022<\/td>\n<\/tr>\n
[1] FS specification applies to the span of the module range. Accuracy includes one-year stability, except for DP025 to DP10 modules.<\/td>\n<\/tr>\n
[2] Accuracy is a 6 months spec, 1-year long-term drift is 0.2%FS.<\/td>\n<\/tr>\n
[3] Accuracy is a 6 months spec, 1-year long-term drift is 0.1%FS<\/td>\n<\/tr>\n
[4] Accuracy is a 6 months spec, 1-year long-term drift is 0.05%FS.<\/td>\n<\/tr>\n
[5] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring in absolute mode. Applicable only for use with the ADT760-MA<\/td>\n<\/tr>\n
[6] The low module pressure range may be outside the pressure range of the calibrator<\/td>\n<\/tr>\n
* ADT155 Pressure Modules are calibrated in psi and inH2O.<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n

 <\/p>\n<\/div>\n

\n
\n\n\n\n\n\n\n\n\n\n\n\n\n
Electrical Specification<\/strong><\/th>\n<\/tr>\n
\u00a0<\/strong><\/th>\nRange<\/strong><\/th>\nResolution<\/strong><\/th>\nAccuracy<\/strong><\/th>\nNote<\/strong><\/th>\n<\/tr>\n<\/thead>\n
mA Measure<\/strong><\/td>\n\u00b130 mA<\/td>\n0.0001 mA<\/td>\n0.01%RD + 1.5 uA<\/td>\nImpedance<10\u2126<\/td>\n<\/tr>\n
V Measure<\/strong><\/td>\n\u00b130 V<\/td>\n0.0001 V<\/td>\n0.01%RD + 1.5 mV<\/td>\nImpedance >1M\u2126<\/td>\n<\/tr>\n
mA Source<\/strong><\/td>\n24 mA<\/td>\n0.001 mA<\/td>\n0.01%RD + 1.2 uA<\/td>\n20 mA @ 1K<\/td>\n<\/tr>\n
Loop Power Source<\/strong><\/td>\n24 V<\/td>\nN\/A<\/td>\n\u00b11 V<\/td>\n50 mA (Max Loading)<\/td>\n<\/tr>\n
Pressure Switch<\/strong><\/td>\nOpen, close. Support for mechanical switches and NPN\/PNP digital switches.<\/td>\n<\/tr>\n
Temperature Compensation<\/strong><\/td>\n47\u00b0F to 100\u00b0F (8\u00b0C to 38\u00b0C)<\/td>\n<\/tr>\n
Temperature Coefficient<\/strong><\/td>\n< \u00b1 ( 0.001%RD + 0.001%FS ) \/ \u00b0C outside of 8\u00b0C to 38\u00b0C<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/div>\n

 <\/p>\n<\/div>\n

\n
\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n\n
General Specifications<\/strong><\/th>\n<\/tr>\n
Specification<\/strong><\/th>\nDescription<\/strong><\/th>\n<\/tr>\n<\/thead>\n
Channels<\/strong><\/td>\nFour total: one electrical, one internal pressure, two external pressure (measure only)<\/td>\n<\/tr>\n
Enclosure IP Rating<\/strong><\/td>\nIP52 water and dust proof<\/td>\n<\/tr>\n
Battery<\/strong><\/td>\nRechargeable Li-Ion battery, typically 10 hours of operation, recharges in less than 4 hours<\/td>\n<\/tr>\n
Display<\/strong><\/td>\nColor 800 x 480 TFT 5-inch touch screen<\/td>\n<\/tr>\n
Communications<\/strong><\/td>\nUSB, WiFi<\/td>\n<\/tr>\n
Weight<\/strong><\/td>\n<4 lbs (<1.8 kg)<\/td>\n<\/tr>\n
Size<\/strong><\/td>\n9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm)<\/td>\n<\/tr>\n
Certification<\/strong><\/td>\nISO17025 accredited certificate with data included<\/td>\n<\/tr>\n
HART Communications<\/strong><\/td>\nOptional (ADT760-X-DL model)<\/td>\n<\/tr>\n
Data Logging<\/strong><\/td>\nOptional (ADT760-X-DL model), up to 1,000,000 readings (date and time stamped)<\/td>\n<\/tr>\n
Task Documentation<\/strong><\/td>\nOptional (ADT760-X-DL model) up to 1000 tasks<\/td>\n<\/tr>\n
Automation Functions<\/strong><\/td>\nSwitch test, auto step, leak test<\/td>\n<\/tr>\n
Misuse Protection<\/strong><\/td>\nUp to 30 V on any two sockets<\/td>\n<\/tr>\n
Multi Lingual Interface<\/strong><\/td>\nEnglish, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian<\/td>\n<\/tr>\n
Pump Life<\/strong><\/td>\n>500,000 cycles<\/td>\n<\/tr>\n
Power<\/strong><\/td>\nRechargeable Li-Ion battery, external power: 110\/220 V power adapter 10 V<\/td>\n<\/tr>\n
Environment Specifications<\/strong><\/td>\nOperation: 32\u00b0F to 122\u00b0F (0\u00b0C to 50\u00b0C), 0-90% RH, less than 3,000 meters Compensated temperature: 32\u00b0F to 122\u00b0F (0\u00b0C to 50\u00b0C) Storage temperature: -4\u00b0F to 158\u00b0F (-20\u00b0C to 70\u00b0C)<\/td>\n<\/tr>\n
Vibration and Shock<\/strong><\/td>\nVibration: 4g (20 to 2,000 Hz) Shock: 8g, 1 meter drop test<\/td>\n<\/tr>\n
Compliance<\/strong><\/td>\nCE<\/td>\n<\/tr>\n
Software<\/strong><\/td>\nACal, PCal, Land, LogII<\/td>\n<\/tr>\n
Warranty<\/strong><\/td>\n1 year<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n