為細胞培養(yǎng)提供出色的加熱和保護
增加自動化細胞培養(yǎng)培養(yǎng)的產(chǎn)量,而不占用寶貴的臺面空間。Inheco單板培養(yǎng)箱小巧靈活,可快速達到工作溫度。
恒溫箱提供受保護的加熱環(huán)境,細胞樣本可以在最佳條件下生長,避免光照和污染物。
節(jié)省空間
Inheco孵化器位于甲板外或甲板旁邊,以節(jié)省您的空間。它通過自己的內(nèi)部控制器進行調(diào)節(jié),實現(xiàn)甲板上的自動化應用。其自動打開和關閉的門將樣品板移出,即使只有該部分伸入液體處理平臺的工作區(qū)域,也能實現(xiàn)甲板上的自動化應用。
快速準確
Inheco單板培養(yǎng)箱只需10分鐘即可達到工作溫度。為了向您提供高度精確的分析,我們的培養(yǎng)箱最高工作溫度為80°C,溫度精度為0.3 K,均勻性為0.5 K。
多才多藝的
微孔板有兩種型號可供選擇:僅從底部加熱,適用于有蓋微孔板;深孔:從各個方向加熱,以避免冷凝,適用于樣品瓶和容器。為了獲得更大的靈活性,可以將兩個或更多版本堆疊在一起。所有這些都在149毫米x 268毫米的工作站空間內(nèi)。放置在最適合您的設置的位置–您可以將保育箱放置在不同的方向,以便于插入和取出孔板,并且無需從抽屜中取出孔板即可完成移液。
易于集成
API可用,并且已經(jīng)與主要液體處理系統(tǒng)集成。對于堆疊設備,設備是單獨和獨立控制的(主從原則)。
Technical Specifications Incubator
Incubator type |
Single Plate Incubator MP |
Single Plate Incubator DWP |
Input voltage |
24 Vdc |
Input power |
Max 30 W |
Max 45 W |
Maximum numbers of devices per pow-er supply (2400121) |
6 |
4 |
Temperature range |
Room temperature +5 K up to +80 °C (+175 °F) |
Tolerable relative humidity ambient |
10-80% relative at +20 °C (+68 °F) up to +30 °C (+86 °F) |
Maximum load of the shaker platform |
N.A. |
Shaker frequency range |
N.A. |
Maximum shaking amplitude (difference between the maxima) |
N.A. |
Shaking direction |
N.A. |
Dimension in YModule?w connector cable |
295.5 mm |
Dimension in YModule?wo connector cable |
268.5 mm |
Dimension in XModule |
149 mm |
Dimension in ZModule?wo pin |
58 mm |
104 mm |
Open drawer incl. door YDrawer |
193.5 mm |
223.5 mm |
Dimension of the thermal chamber in ZChamber |
18 mm |
50 mm |
Weight |
2.5 kg |
3.4 kg |
Maximum number of devices in one stack due to interaction during shaking and power limitation of power supply |
5 |
4 |
Drawer specifications |
Accuracy of the end position of the drawer in x, y and z (drawer open); single unit, all parts at specified ambient temperature |
±0,30 mm |
Maximum drawer load |
0,25 kg |
Open drawer resistance for a force in Z-axis without plastic deformation |
10 N |
Environmental conditions |
Tolerable relative humidity: |
Operation |
10-80% relative at+20 °C (+68 °F) up to +30 °C (+86 °F) |
Transportation and storage |
-10 °C to +60 °C (+14 °F to +140 °F) |
Temperature: |
Operation |
+20 °C to + 30 °C (+68 °F to +86 °F) |
Transportation and storage |
-10 °C to + 60 °C (+14 °F) to 140 °F) |
Shaker functionality Incubator Shaker MP / DWP
Parameter |
Value |
Amplitude range |
0.0 – 3.0 mm * |
Speed range |
6.6 – 30.0 Hz * |
Speed precision |
±2 Hz |
Load |
40 – 250 g * (weight of plate and liquid and adapter) |
Supervision |
Speed and amplitude (closed loop controller) |
Shaker movement |
North – South (linear)
East – West (linear)
Round (orbital)
“Eight” (this is possible but not recommended because it might damage the shaker) |
Shaker movement |
Amplitude accuracy** ±20% (of the target amplitude)
Phase shift accuracy** ±11 °
** Validated for the Heating Plate movement / average value over 10 cycles |
Single Axis (North – South and East – West) Shaker Load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Maximum stack |
10 |
X = 1.0 to 2.2 or Y = 1.0 to 2.2 |
X = 1,0 to 1.9 or Y = 1.0 to 1.9 |
15 |
X = 1.0 to 2.7 or Y = 1.0 to 2.7 |
X = 1.0 to 2.4 or Y = 1,0 to 2.4 |
20 |
X = 1,0 to 1.2 or Y = 1.0 to 1.2 |
X = 1.0 to 1.2 or Y= 1.0 to 1.2 |
Round (Phase shift = 90 °) |
Shaker load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Maximum stack |
10 |
X = 1.0 to 2.2
Y = 1.0 to 2.2 |
X = 1,0 to 1.9
Y = 1.0 to 1.9 |
15 |
X = 1.0 to 2.7
Y = 1.0 to 2.7 |
X = 1.0 to 2.4
Y = 1.0 to 2.4 |
20 |
X = 1.0 to 1.2
Y = 1.0 to 1.2 |
X = 1.0 to 1.2
Y = 1.0 to 1.2 |
?Eight“ (Phase shift = 0°) |
Shaker load: 100 g |
Shaker |
Shaker amplitude [mm] |
Speed [Hz] |
Single device |
Triple stack |
X = 12 / Y = 24 |
X = 1.0 to 2.4 / Y = 1.0 to 2.4 |
X = 1.0 to 2.2 / Y = 1.0 to 2.2 |
X = 15 / Y = 30 |
X = 1.0 / Y = 1.0 |
X = 1.0 / Y = 1.0 |
|
Thermal functionality for all Incubators
Parameter |
Value |
Temperature range |
Room temperature plus 5 K to +80 °C [+176 °F] |
Accuracy absolute (@ ambient temperature stability ±1 K) |
The accuracy is defined as the difference between the target temperature of the heating plate and the average of the measured temperature on the heating plate. The heating plate temperature is measured at 5 positions on the heating plate according to the Inheco measurement head PM-0105. The 5 measurement positions are according to the Plate positions A1, H1, E6, A12 and H12.
Target temp. validation tool |
Accuracy [°C] for all incubator |
+37 °C [+98,6 °F] |
±1.0 |
+60 °C [+140 °F] |
±1.5 |
+80 °C [+176 °F] |
±2.0 |
|
Uniformity measured in well |
Uniformity is defined by the maximum difference between lowest and highest temperature according to a Greiner 96 Flat Bottom Plate (postitions A1, A12, E7, H12)
Target temp. validation tool |
Uniformity [°C] for non shaking devices |
Uniformity [°C] for shaking devices |
+37 °C [+98,6 °F] |
≤ 0.8 |
≤ 0.8 |
|
Evaporation in unsealed Micro Plates (can be reduced by using lid or cover) |
Incubator?non?shaker temperature +37 °C [+98,6 °F],?max. 25% loss?of liquid in?4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level |
Incubator shaker temperature +37 °C [+98,6 °F],?max. 32% loss?of liquid in?4 h, deionized water, room temperature 24-26 °C [75,2-78,8 °F], rel. humidity 40-50%, altitude at 410 m over Sea level |
|
Heat-up time at start temperature +20°C [+68°F] with target temperature +37°C [+98°F] |
Max. 30 min; Volume 280 μl per well. * |
Max. 100 min; Volume 1.5 ml per well.?** |
Max. 30 min; Volume 280 μl per well.?* |
Max. 120 min; Volume 1.5 ml per well.?** |
|
* The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1)?
Greiner 96 Flat Bottom Plate
** The target temperature is reached when all sensors are in the tolerance band of 37 °C ±1°C [98,6 °F ±3,6 °F] Measured with: 1)?NUNC 96 Deep Well Plate