The T92+ wavelength range is 190–900nm and the instrument has a User Defined Variable Bandpass from 0.1nm – 5.0nm. The photo metric range is -4 to +4 Abs.
T92波長范圍為190~900 nm,儀器具有用戶定義的0.1nm~5.0nm的可變帶通。攝影測量范圍是-4到4 Abs
具有全息光柵的Czerny-Turner單色器使雜散光保持最小并提供極好的光學分辨率。
光電倍增管作為檢測器的使用提供了卓越的靈敏度
T92的真雙光束光學系統與高效且成熟可靠的電子控制系統相耦合,確保了高穩定性和低背景噪聲
Optical System | Czerny-Turner Monochromator with high resolution grating |
Spectral Band width | 0.1 – 5nm user definable in steps of 0.1nm |
Working Mode | PC Mode with USB connection. |
Software Support | UV/Win-Pro Software |
Wavelength range | 190 – 900nm |
Wavelength Accuracy | ± 0.3nm (Automatic Wavelength Correction) |
Wavelength Reproducibility | 0.1nm |
Light Source | Automatic interchange (selectable within the working range of light source) |
Stray Light | ≤ 0.01% T (220nm) |
Photometric mode | Transmission, Absorption, Reflectance, Energy & Concentration |
Photometric Range | -4.0 to 4.0 Abs |
Photometric Accuracy | ± 0.002Abs (0-0.5A)± 0.004Abs (0.5 -1.0A) ± 0.3%T (0-100%T) |
Photometric Reproducibility | 0.001Abs (0 – 0.5A) |
Baseline Flatness | ± 0.001Abs |
Baseline Stability | 0.0004Abs/Hr (500nm, after pre heat) |
Photometric Noise | ± 0.001Abs (500nm,0Abs, 2nm Spectral Bandwidth) |