Vortex probe VA40 with sensor head diagonal dimension 40 mm and transducer UVA integrated in the connection housing. Probe material: stainless steel. Working temperature range up to +100 or +180 °C.
Suitable for fixed measuring of actual flow velocity v [m/s] and actual flow rate V/t [m3/h] in clean as well as condensate and particulate-laden gases.
Signal evaluation with integrated transducer UVA.
according to drawing ZG7 | |
Measuring range: | 0.5 … 40 m/s |
Working pressure range: | up to 3 bar / 300 kPa overpressure |
Working temperature range: | -20 … +100 / +180 °C |
SIL approval: | SIL2 |
Medium: | gases |
Displayed meaurement categories: | flow velocity, flow rate, volume |
IP? protection: | IP65 (connection housing) |
Ex protection: | ATEX category 3G and 3D (zone 2 and 22) (optional) |
Sensor outputs: | analog output 4-20 mA, limit value, quantity pulse, RS232 interface |