H?ntzsch GmbH 公司Probe ZS25 ... ZG5探頭

Probe ZS25 … ZG5 探頭

according to drawing ZG 5
Measuring range: 0.3 … 120 m/s in air/gases,
0.03 … 10 m/s in water/liquids
Working pressure range: up to 10 bar / 1 MPa overpressure
Working temperature range: -40 … +500 °C
Medium: gases / liquids
Displayed meaurement categories: flow velocity, flow rate, volume
IP protection: IP65 (connection housing)
Ex protection: ATEX category 1/2G and category 1/2D (zone 0/1, zone 20/21)
Sensor outputs: analog output 4-20 mA, limit value, quantity pulse, HART communication for configuration