Lambda-messtechnik防護(hù)熱板法導(dǎo)熱系數(shù)儀EP500e、熱導(dǎo)率測(cè)試儀EP500e
技術(shù)規(guī)格
Parameter | Version A | Version C |
Real?Measurement Range?is the?Thermal Transmittance Resistance?of the inserted sample (correlated with the measurement of the electrical heating power of the measuring heater): | R =0.1 … 14 m2K/W | R = 0.025 … 14 m2K/W |
corresponding?Thermal Conductivity?(ca.): | λ = 0 … 0.5 W/(m·K) | λ = 0 … 3 W/(m·K) |
Measuring Area/Zone: | 200 x 200 mm2 | 150 x 150 mm2 |
Sample Thickness: | 10 … 120 mm (optional 4 … 200 mm) | |
Sample Dimensions: |
| |
Reproducibility: | < 0.5 % (mostly < 0,2 %) | |
Accuracy: | < 1.0 % (mostly < 0,7 %) | |
Measurement Temperature Range?(Standard): | 10 … 40 °C (optional -10 … 50 °C) | |
Temperature Difference Range?(Standard): | 5 … 15 K (optional 20 K) |