t2 Sport/EVO 支架 光學支架 精密溫控樣品池支架

Peltier-driven temperature control for the Evolution 300 and 600 Spectrophotometers ? Evolution300和600分光光度計的 peltir 驅動溫度控制

The t2 Sport/EVO is an inexpensive, compact cuvette holder. Use it for temperature ranges of about -15 °C to 110 °C. It is ideal for most biological studies to observe absorbance spectra at precisely maintained temperatures. The t2 Sport/EVO mounts in place of the forward sample cuvette holder with utilities conveniently routed under the front panel of the spectrophotometer. Operate the TC 1/Single Temperature Controller using the menu button or control via optional program T-App.
t2 sportyevo 是一個價格低廉, 緊湊的 cuvette 支架。在-15°c 至110°c 的溫度范圍內使用。它是大多數生物研究的理想選擇, 在精確保持的溫度下觀察吸收光譜。t2 sporty-evo 安裝在向前樣品杯架的位置, 在分光光度計的前面板下方便地布線。使用菜單按鈕操作 tc 半溫度控制器, 或通過可選程序 t-app 進行控制。

Product Features

Rapid, precise Peltier temperature control快速、精確的 peltier 溫度控制
Variable speed magnetic stirring?變速磁攪拌
Dry gas purge for work at low temperature?低溫工作用干氣凈化
Base and post for mounting in the spectrophotometer?分光光度計中安裝的基礎和立柱
Cuvette lifter for removal of cuvette?用于去除立方的立方提升機
Probe input for sample temperature?樣品溫度的探頭輸入
NIST-traceable temperature calibration
nist 可追溯溫度校準

技術資料:

Number of cuvettes controlled:1
Typical use:Absorbance
Standard cuvette size (outside):12.5 mm x 12.5 mm
Optical ports per cuvette:2
Optical port dimensions:12 mm high x 10 mm wide
Cuvette z-height:8.5 mm or 15 mm
Windowed jacket available:No
Factory set temperature range:-40 °C to +110 °C
Temperature range easily achievable:-15 °C to +110 °C
Temperature precision:±0.02 °C
Temperature accuracy:Better than ±0.15 °C from 0 ?C to +80 ?C
Accuracy full range:Better than ±0.3 ?C from ‐15 ?C to +110 ?C
Temperature reproducibility:Better than ±0.07 °C
Thermistor probes accepted:Series 400 or Series 500
Magnetic stirring motor type:DC
Magnetic stirring speed range:100-2500 rpm
Default stirring speed:1200 rpm
Stirring speed best performance:200-1800 rpm

Basic Package
t2 Sport – Temperature-controlled cuvette holder
TC 1/Single – Temperature Controller
Base plate and post (standard z-height = 15 mm)
BATH 10 – Circulator
NIST-traceable calibration and performance data
Accessories – Magnetic stir bar, tubing and cables

Accessories
T-App
Water Circulators
Cuvette Adapters
Thermistor Probes