Tyne’s Tritium Controller is the main data processing centre for many of our tritium monitoring configurations. It is suitable for in-line monitoring of process systems, glove box monitoring, stack monitoring, room air monitoring, and other applications such as with Tyne’s tritium-in-breath analyzer.?Tyne的氚控制器是我們許多氚監測配置的主要數據處理中心。適用于過程系統的在線監測、手套箱監控、棧監控、室內空氣監測等應用, 如與Tyne的氚呼吸分析儀。
It can be configured for a variety of tritium monitoring requirements, including inputs from one or two ion chambers of different sizes and sensitivities (see Tyne’s 1000cc ion chamber and 10cc ion chamber), multiple alarm configurations, and multiple range configurations.它可以配置為各種氚監測要求, 包括來自一個或兩個不同大小和靈敏度的離子室的輸入 (見Tyne的1000cc 離子室和10cc 離子室), 多個報警配置, 和多個范圍配置。
特征
大型平板觸摸屏
最多可控制兩個探測器
每通道雙報警, 本地和遠程
一個觸摸自動零點
4-20 mA、0-10V 或 RS232 輸出
平滑的汽車范圍橫跨9年
Specifications
Note: Specifications depend on configuration and ion chamber selection.
Detection Sensitivity with 1000 cc Detector | 1 μCi/m3 |
Detection Sensitivity with 10 cc Detector | 0.1 mCi/m3 |
Instrument Response Time | less than 10 s |
Detection Range with 1000 cc Detector | 0 to 10 Ci/m3 |
Detection Range with 10 cc Detector | 0 to pure tritium |
Flow Rates with 1000 cc Detector | 1000 cc/s recommended, 2000 cc/sec max |
Flow Rates with 10 cc Detector | 1000 cc/s |
Operating Pressure | Vacuum to 150 psig |
Operating Temperature (ion chambers) | 15 °C to 200 °C |
Operating Temperature (controller) | 0°C to 50°C |
Operating Humidity | 0% – 95% Relative Humidity |
Leak Tightness | less than 1×10-9 atm-cc/s He |
Temperature Offset | Under 1 μCi/°C |
Gamma Offset | 0.1 Ci/m3 per 1 R/h gamma fieldLinear response from 10-3 R/h to 102 R/h |
Gamma Compensation | Subtract with 2nd detector head |
Background Compensation | “One touch” Auto Zero |
Tritium Recovery | Chamber bakeable up to 350°C |
Tritium Wetted Parts | 316L Stainless Steel, High density Ceramic |
Fittings | VCR8 |
Controller Placement | Up to 100m from detector head |
Analogue Output | 0 – 10 V |
Current Output | 4 – 20 mA |
Digital Output | RS232 |
Range Indication | 3 digital output lines |
Process Alarm | Two alarm set points per channel |
Process Control on Alarm | Alarm Relay, 30 VAC/DC, dry contact relay |
Compliance | DOE Tritium Monitor Standard Rev 4 June 1999, ISO 9001:2000 |
Power Requirements | 1A, 120 VAC, 50/60 Hz1/2A, 220 VAC 50/60 Hz |
Dimensions of Controller | Standard 19″ rack mount x 5″ Height |
Dimensions of 1000 cc Detector | 4.5″ diameter x 15″ length |
Dimensions of 10 cc Detector | 3″ diameter x 15″ length |
Power | 120 VAC 1A or 220 VAC 1/2A 50/60Hz |