7009 Switching Chamber Tritium-in-Air-Monitor 開關室測氚儀

Tyne’s tritium in air monitor is used to determine the total tritium in air based on a continuous air sample drawn through a flexible tube that can be located up to 100 feet away. The tritium is measured using very accurate ion chambers. Tyne在空氣監測儀中的氚是用來測定空氣中的總氚的基礎上,通過一個可位于100英尺以外的柔性管的連續空氣樣品。氚是用非常精確的離子室來測量的。

Specifications

Range using Tyne controller, preamp, 1000cc ion chamber 0 through 1000 Ci/m3
Max sensitivity 1 μCi/m3
Tritium recovery Bakeable to 250°C, washable with demineralized water
Controller display 108mm x 58mm touch screen
Accuracy 10% of monitoring range
Power 120V 50/60 Hz
Output Signal 4-20 mA or 0-10V selectable
Response time using Tyne controller, preamp, ion chamber 20 secs for measurements to 80 μCi/m3; 3 seconds for measurements between 80 μCi/m3 and 1 mCi/m3.
Ion Chambers 2 x 1000cc enclosed volume type
Ambient Temperature 0°C to 50°C
Relative Humidity 0 to 95%
Temperature Offset using Tyne controller, preamp, ion chamber Less than 1 μCi/°C (can be computer compensated)
Max Flow rate 2,000 cc/s
Dimensions of Controller 381 mm x 483 mm x 140 mm (19″ Rack Mount)
Dimensions of Area Tritium Monitor 381 mm x 483 mm x 140 mm (19″ Rack Mount)
Electrical Safety EMI Shielded
QA ISO 9001:2000