7043 Portable Tritium-in-Air-Monitor 便攜式測氚儀

Compensation for Gamma background is crucial in tritium measurement, since a gamma field of 1mR/Hr will generate 500 times the ionization generated in 1μCi/m3 of tritium. Tyne designed the four, 250 cc ion –chambers in a cruciform with each chamber connected to an ultra sensitive electrometer amplifier and filter. All signals pass through the ADC converter, and the micro-processor calculates and displays the tritium value. To cover high ranges of tritium a separate small ion chamber is used. Measurements such as flow from the solid-state flowmeter and chamber temperature are also monitored by the micro-processor.?對γ背景的補償是氚測量的關鍵,因為1 mR/HR的伽瑪場產生的電離量是1 mI/m氚電離的500倍。TYNE設計了四,250毫升離子室在一個十字形,每個室連接到一個超靈敏的靜電計放大器和濾波器。所有信號通過ADC轉換器,微處理器計算并顯示氚值.為了覆蓋高范圍的氚,使用了一個單獨的小離子室。測量結果,如來自固態流量計的流量和室內溫度,也由微處理器進行監測。

Specifications

Sensitivity 1μCi/m3 ( 0.05 DAC)
Accuracy ±3μCi/m3 from 1 to 100μCi/m3, ±10% from 100μCi/m3 to 20,000 μCi/m3
Range 1 μCi/m3 to 199,999 μCi/m3
Detector 4 matched chambers – Two for measuring, two for compensation
Measuring chamber effective volume 500 cc (2 measuring chambers of 250cc each)
Flow rate 1.0 L/m
Zero stability ±1μCi/m3 immediately after unit is powered on
Background cancellation Less than 10% of the reading change in 20 mR/hr field
Radon compensation Included in the software
Ion trap Built into each ion chamber
Dust filter Built into the unit.
Noble gas cancellation Ports for air inlet/outlet of measuring chambers and compensation chambers are accessible on surface plate so silica gel drier can be placed between the chambers to compensate for presence of noble gas.
Tritium Discrimination Instrument can discriminate in a field of HTO and elemental H3 using the silica gel drier.
Purge/Decontamination All 4 Chambers can be heated up by a cartridge heater controlled by unit
Display Graphic touchscreen LCD display. Unit displays the values(tritium and Gamma) both in numerical format and graphical trend.
Communication Can be configured to RS232(Short distance) or RS485(Long distance).
Analog output 0-5 V Linear or logarithmic voltage output.
Power supply 2.5 V- 5V external power supply, or 3 C-cell battery. Total power consumption is 150mA.
Gamma measurement Gamma is measured directly by the unit. Gamma measuring range is from 0.1mR/h to 10R/h
Alarm Tritium alarms, gamma alarms, low air flow alarms will all be shown by red LED light and by an audible device.
Data logging The tritium and gamma values are logged every minute with current time and date stamp
Size and weight 8″(L)x5″(H)x5″(W), 7 lbs.
Drier Note that the silica gel drier is not part of the standard equipment, but can be supplied as an extra.
Response Time 30 seconds at 90% of signal