Tyne Engineering’s new Model 7045 Room Tritium In-Air Monitor is a wall or rackmount version of our popular Model 7043 Portable Tritium Monitor. The 7045 is intended for room monitoring for both tritium and gamma. Tyne的新型7045室氚空氣監測器是我們流行的型號7043便攜式氚監測器的墻或機架式版本。7045的目的是為氚和伽瑪的房間監測。
Specifications
Measuring chamber effective volume | 1.5 L, (2 X 0.75L) |
Pump | Positive displacement gas pump: 12 VDC, flow rate can be up to 11L/min, continuous pressure 8 psig. |
Tritium measuring range | 0.2μCi/m3 (7,400Bq/ m3) to 1Ci/m3 |
4 chambers | cruciform configuration: two for measuring and two for compensation |
Radiation measured | measuring total tritium, HTO, gamma and alpha |
Sensitivity | 0.2μCi/m3 (0.01 DAC) |
Accuracy | 0.4μCi/m3 from 0 to 100μCi/m3; ±10% from 100μCi/m to 1Ci/m3; |
Zero stability | ±0.4μCi/m3; |
Gamma measurement | Gamma is measured simultaneously with tritium. Measuring range is from 0.1mR/h to 10 R/hr; |
Radon compensation and tritium discrimination | Instrument can provide radon compensation and discriminate in a field of HTO and elemental H3 using the silica gel drier. |
Gamma background compensation | less than ±10% or 5mCi/m3 reading change whichever is greater response to 20mR/Hr; |
Ion trap | built into each ion chamber; |
Flow rate | 0 to 11 L/min |
Response time | 2 to 5s with 0-15 m air hose, within 20s with hose from 100 to 150m at 90% of signal |
Stability | : ±3% per 24 hours, ±10% per 30 days; |
Analog Output | 0-3V linear. |
Alarms | Tritium, gamma, air flow, detector failure, micro processor alarms in field as well as on optional remote display system through analog output (4 to 20mA) |
Alarm Response | < 30S with 2-150m hose; |
Purge decontamination | Chambers can be heated by a cartridge heater controlled by the unit to assist in purging, |
Condensation Prevention | With heater cartridge temperature can be raised above ambient to prevent condensation in instrument |
Communication port | RS232 (short distance) and RS485-bus (long distance); |
Data Logging frequency | The tritium value can be logged every minute and can be time and date stamped. (Gamma is not logged) |
Displays | One (1) TFT LCD large touch screen display (size 2.8’’), displays the values of tritium, gamma in both numerical and graphical format; one LED screen ?’’ letters, 7 letter screen. |
Power supply | 220VAC (±10%); Power consumption < 1A |
Drier | silica gel cylinder is provided with the unit; |
Temperature operation range | 0 to 50oC degree; |
Humidity operation range | 0-90% Non-condensable |
Size | Approx 9″High x 12″Deep x 19″ Wide; |
Weight | Approx.: 55lb; |
Mounting | Unit is designed for table-top, 19” rack, or wall-mounting (bracket optional) |