The Single Range Tritium Signal Processor is used to measure the amount of tritium directly in line of a process loop in real time and transmit a 0 – 10V output signal The unique small footprint detector head has the ability to handle up to 150 psig. It is leaktight and can operate at elevated temperatures. Up to 10 transmitters may be powered by one low cost power supply making this the most cost effective solution for a process with many detection locations.
采用單量程氚信號處理器實時測量過程回路中的氚量, 并傳輸 0-10V 輸出信號, 獨特的小足跡檢測頭具有處理多達 150 psig 的能力。它是 leaktight 的, 可以在高溫下操作。多達10臺發(fā)射機可能由一個低成本電源供電, 這是一個具有許多檢測位置的過程中最經(jīng)濟劃算的解決方案。
特征
直接內(nèi)工藝線安裝
低靈敏度為1μCi/立方米
150 Psig 壓力容器
操作溫度高于100°c
1×10-9 泄漏
0-10 V 輸出, 用于集成到數(shù)據(jù)系統(tǒng)中
用戶可選擇的規(guī)模多達1萬個單位
低成本
Specifications
Detection Sensitivity with 1000 cc Detector | 1 μCi/m3 |
Detection Sensitivity with 10 cc Detector | 1mCi/m3 |
Detection Range with 1000 cc Detector | 0 to 10 Ci/m3 |
Detection Range with 10 cc Detector | 0 to pure tritium |
Measurement Steps with 1000 cc Detector | 10,000 steps |
Measurement Steps with 10 cc Detector | 10,000 steps |
Flow Rates | 2000 cc/s |
Flow Rates | 1000 cc/s |
Operating Pressure | Vacuum to 150 psig |
Operating Temperature | 15 °C to 200 °C |
Operating Humidity | 0% – 95% Relative Humidity |
Leak Tightness | less than1x10-9 atm-cc/s He |
Temperature Offset | Under 1 mCi/°C |
Gamma Offset | 0.1 Ci/m3 per 1 R/h gamma fieldLinear response from 10-3 R/h to 102 R/h |
Tritium Recovery | Chamber bakeable up to 350°C |
Tritium Wetted Parts | 316L Stainless Steel, High density Ceramic |
Fittings | VCR8 |
Instrument Response Time | less than 10 s |
Analogue Output | 0 – 10 V |
Compliance | DOE Tritium Monitor Standard Rev 4 June 1999 ISO 9001 |
Power Requirements | 135 VDC floating from ground, 0.1 A |
Dimensions with 1000 cc Detector | 4.5″ diameter x 15″ length |
Dimensions with 10 cc Detector | 3″ diameter x 15″ length |