MS-50GT HIGH SPEED PRODUCTION CONSOLE HELIUM LEAK DETECTOR 高速生產控制臺 氦檢漏儀

MS-50GT HIGH SPEED PRODUCTION CONSOLE HELIUM LEAK DETECTOR 高速生產控制臺 氦檢漏儀

The MS-50GT is a highly technical system that incorporates many sophisticated technologies. In view of this, every effort has been made to automate the unit’s operation and maintenance procedures. The MS-50GT is an automated high-speed production console leak detector featuring three flow modes which provide flexibility and maximum productivity for application-specific leak testing.

It also meets the stringent requirements for robustness, reliability and ease of maintenance that are essential in any industrial environment..

MS-50GT是一個高度技術化的系統,融合了許多尖端技術,該裝置的操作和維護程序自動化。

MS-50GT是一款自動化高速生產控制臺泄漏檢測器,具有三種流動模式,可為特定應用的泄漏測試提供靈活性和最大生產力。
它還滿足在任何工業環境中都至關重要的堅固性、可靠性和易于維護的嚴格要求

MS-50GT:?APPLICATIONS

  • AUTOMOTIVE

    • Exhaust System
    • Cooling Circuit
    • Engine
    • Battery
    • Oil Pan
  • AIR CONDITIONING

    • Condenser Coils
    • Evaporator Coils
    • Compressor Seals
    • Refrigerant Tubing
    • Valve and Joint
  • MEDICAL

    • IV Bag
    • Implantable Device
    • Syringe and Vial
    • Medical Pouch
    • Packaging
  • NUCLEAR

    • Fuel Rod
    • Coolant System
    • Containment Vessel
    • Storage Container
    • Glove Box
  • AEROSPACE

    • Fuel System
    • Hydraulic System
    • Oxygen System
    • Pressurized Cabin
    • Cooling System

     

    MS-50GT:?FEATURES

    Characteristics

    • Sensitivity, vacuum*:?3.0×10-11 (atm-cc/sec)
    • Sensitivity, sniffing:?1.0×10-08 (atm-cc/sec)
    • Leak Measurement Range:?3.0×10-111 to 10
    • Leak Ranging:?Auto/Manual
    • Startup Time:?Less than 3 minutes
    • Response Time *:?<2 direct <2.5 reverse
    • Resolution for 4He*:?14 at mass 4
    • Gross Test Time (with UFT Option):?– 3 (100 Torr)(seconds)
    • Fine Test Time (with UFT Option):?– 10 (25 mTorr, 10^-10 atm-cc/sec)
    • Mass Spectrometer Tube:?180° deflection; dual-magnetic sector
    • Auto-tuning
    • Auto-calibration:?Internal; External – auto
    • Auto-zeroing:?Several zeroing functions
      modes for background suppression
    • Reject set-point
    • Built-in calibrated leak:?10-08 range
      NIST traceable: Yes
    • Temperature compensated
    • Scan/Descan
    • Filaments:?2 Thoria Coated Irridum
    • Testable Masses:?3+4
    • Background Compensation:
    • Automatic
      Maximum Operating Pressure:?100 mTorr, 10 Torr reverse
    • Display/Interface:?10 (25.4) LCD flat panel display, capacitive
      touch-screen / Graphical
    • User?Interface (GUI?Units:?cc/second, mbar*l/second, Pa*M3/second
    • RS-232C
    • Software:?50 programmable test recipes; Data Link optional

     

    Measurement Characteristics
    Sensitivity VacuumSensitivity SniffingStartup Time
    3.0×10-11
    (atm-cc/sec)
    10-3~10-5
    Atm-cc/second
    Less than 3 minutes
    MS-50GT:?CHARACTERISTICS

    Available Models

    • Dual Port
    • Ultrafast (UFT)
    • Oil-based
    • Dry Pump
    • High Sensitivity
    • Argon Gas

     

    Technical Specifications
    Power115/50-60/30; 230-240/50-60/15
    100/50-60/30 V/Hz/A
    Dimensions(W x H x D): 25.75 x 30.625 x 35.25
    (65.40 x 77.79 x 89.54)
    Weight475 lbs,? 215.45 kg
    Vacuum ModeDirect, reverse or combination, gross testing
    Roughing Pump16 direct drive (cfm)
    Test Port Connection1.125 (inches)
    High Vacuum Pump53 L/sec
    Vacuum GaughingPirani: Test Port and Foreline
    Ion: Hihg Vacuum
    Cold TrapMS-50GT – N/A
    MS-50GT Dry – N/A
    Test Cup, UFT
    Option UFT(? x D)
    Dimensions (in mm)
    W344*H395*D618